Instrumentation Used in Clinical Gait Studies a Review

Skip Nav Destination

Research Article |

Gait Analysis

Victoria L. Chester, PhD;

Victoria L. Chester, PhD *

From the Kinesthesia of Kinesiology, Constitute of Biomedical Engineering, University of New Brunswick, Fredericton, NB, Canada (VC); Department of Mechanical Engineering, Plant of Biomedical Engineering, University of New Brunswick, Fredericton, NB, Canada (EB); and Department of Mathematics and Statistics, University of New Brunswick, Fredericton, NB, Canada (MT)

* Address correspondence and reprint requests to V. Chester, Faculty of Kinesiology, Constitute of Biomedical Applied science, University of New Brunswick, Fredericton, NB, Canada E3B 5A3 (email: vchester@unb.ca).

Search for other works by this writer on:

Edmund North. Biden, DPhil;

From the Kinesthesia of Kinesiology, Institute of Biomedical Engineering, University of New Brunswick, Fredericton, NB, Canada (VC); Department of Mechanical Engineering, Institute of Biomedical Engineering, University of New Brunswick, Fredericton, NB, Canada (EB); and Department of Mathematics and Statistics, University of New Brunswick, Fredericton, NB, Canada (MT)

Search for other works past this author on:

Maureen Tingley, PhD

From the Kinesthesia of Kinesiology, Institute of Biomedical Engineering, University of New Brunswick, Fredericton, NB, Canada (VC); Section of Mechanical Engineering, Found of Biomedical Engineering science, Academy of New Brunswick, Fredericton, NB, Canada (EB); and Section of Mathematics and Statistics, Academy of New Brunswick, Fredericton, NB, Canada (MT)

Search for other works past this writer on:

Biomed Instrum Technol (2005) 39 (1): 64–74.

Gait analysis, or the study of locomotion, has changed dramatically over the terminal few decades. Advances in reckoner technology and data analysis techniques have contributed greatly to the progress of this field. Gait analysis has go a valuable tool in the clinical setting. The ability to objectively quantify motion is essential to our agreement of normal and abnormal motility patterns and the evaluation of handling effectiveness. This paper volition discuss the various experimental and analytical techniques currently used for performing clinical gait analyses at the University of New Brunswick, Fredericton, New Brunswick, Canada.

Recent advances in computer engineering science and the application of mechanical engineering science to gait analysis have enabled this discipline to develop into a powerful tool in the clinical setting. Clinical gait analysis aims to quantify and to assess the mechanics of walking, and facilitates the identification of deviations from normal movement patterns. The underlying crusade(south) of these abnormalities and their functional consequences are and then adamant in club to provide treatment recommendations. The ability to objectively quantify motility patterns has increased our understanding of normal and pathological gaits and of the effectiveness of specific treatment modalities. As a result, gait analyses tin can pb to savings for the health care system and improved long-term functional condition for patients.1,ii

Clinical gait analysis is a relatively young field that has progressed dramatically over the concluding three decades. During the early 20th century, Inman and colleagues conducted the first comprehensive studies of gait kinematics.3 The measurements of limb positions required the manual digitization of movie images and intensive kinematic computations. This initial application of technology to gait analysis tended to focus on parameters that were appreciable and elementary to measure. For example, clinicians reported spatial parameters such as joint angles, and temporal parameters such as walking velocity.4 Initial attempts to measure more complex and "invisible" kinetic parameters, such every bit external forces and net joint forces, moments, and power were offset conducted during the early 20th century.5,6 By the belatedly 1970s, the appearance of three-dimensional measurement systems led to the improved understanding of these parameters during gait.seven The ability to quantify kinetic parameters is disquisitional, because these parameters are typically being modified with treatment.

Recently, there has been an increment in the use of iii-dimensional quantitative assay in the clinical setting. Many clinical settings now utilise a land-of-the-art motility analysis arrangement equally a routine part of patient assessment, in conjunction with other clinical measures. Many different types of data acquisition systems are bachelor to capture homo motion. However, most labs mostly have three basic systems: 1) a motion capture organization, 2) force plates, and 3) electromyography equipment. These three systems provide the necessary data to estimate kinematic and kinetic gait parameters. The following discussion outlines the various gait data acquisition systems and data analysis techniques used at the University of New Brunswick for the purpose of clinical gait research. In addition, data assay techniques to obtain kinematic and kinetic parameters are addressed.

Motion Capture Systems

One of the nearly popular methods of motility capture for clinical gait assay is the passive optoelectronic method. These are camera-based systems that rails and record the trajectories of lightweight reflective markers located on the patient'due south pare. The Academy of New Brunswick uses a Vicon M-cam motion assay system8 comprised of eight infrared cameras. The cameras are positioned in a task-specific manner around a calibrated measurement volume. As the patient walks back and forth along a marked walkway, the trajectories of the markers are tracked at a sampling frequency of 250 Hz. The three-dimensional locations of the markers are then stereometrically reconstructed from the two-dimensional images of the markers.

During information collection, patients are asked to wear minimal wearable (i.e. bathing suit or underwear) so that the reflective markers can exist adhered directly to the skin in specific anatomical locations. At to the lowest degree three markers are placed on each body segment to obtain the three-dimensional position and orientation of the segment. Markers of various sizes, configurations, and mounting techniques may be used. The University of New Brunswick uses markers with a diameter of 25 mm. Markers mounted on wands are also used on the thigh, shank, and pelvic segments to amplify longitudinal rotations and reduce colinearity. To reconstruct the iii-dimensional locations of the markers, each mark must be visible by at least ii cameras at each instant. In full general, six cameras provide enough redundancy to successfully rail markers on the upper body and both lower limbs simultaneously. The reflective markers are passive and lightweight and do not crave the patient to habiliment electronic cables or power supplies. Therefore, they are considered to have a minimal effect on the natural walking patterns of the patient. The apply of skin-mounted markers tin atomic number 82 to erroneous results if the skin moves relative to the underlying bony landmarks. This is specially evident in obese patients and also when measuring areas of high skin move, such as the patella and the scapula. The magnitude of the skin motion artifact ranges from a few millimeters to 40 mm.9 Care must be taken by the clinician to choose landmarks that are not associated with loftier skin motion (i.due east. greater trochanter).

Modernistic motility capture systems provide rapid marker reconstruction, semiautomatic marker labeling, and playback of the patient's movements in an interactive three-dimensional software environment. The marking labeling process begins by recording the patient while he or she is standing still in the center of the measurement volume. Once the individual markers on this static capture are identified or labeled, the software automatically labels the markers in the proceeding dynamic trials. Compared to manual digitization techniques, automatic tracking greatly reduces data processing time and tracks the centroid of the markers more accurately. Estimates of marking tracking error at the Academy of New Brunswick movement lab more often than not have been less than 2 mm for large volume captures and less than 1 mm for small volume captures.10 Marker occlusion from the photographic camera view and errors due to the proximity of two markers (cantankerous-over fault) are challenges still faced by optoelectronic systems.

Using the three-dimensional position data obtained from motility capture system, it is possible to compute joint angle, temporal-spatial, and kinematic data. Temporal data, such as per centum of fourth dimension in swing stage or walking velocity, can provide of import information near the patient's developmental stage and move stability. Joint bending data provides information on the relative orientation of segments during the gait cycle. Angular parameters are too necessary for the calculation of joint moments and joint power. There are several alternating methods of obtaining these parameters, including accelerometers, radiographic methods, magnetic systems (e.g. Flock of Birds, Ascension Applied science Corporation, Burlington, VT), and active optoelectronic systems, which apply light-emitting diodes instead of passive markers (east.g. Selspot, Innovision Systems Corporation, Columbiaville, MI; Optotrack, Northern Digital Inc., Waterloo, ON, Canada).

Force Plates

While the motility capture system records the patient's body segment movements during walking, force plates simultaneously mensurate the interaction of the foot with the ground. The resultant force generated by this interaction is referred to as the ground reaction forcefulness (GRF). The GRF reflects the dynamic furnishings of the torso (due east.yard. the external forces required to accelerate the centre of mass of the trunk) and back up of the torso weight. The GRF has three components, namely: 1) the vertical force, 2) the anterior-posterior shear, and 3) the medial-lateral shear. Force plates mensurate these three-dimensional forces and the corresponding moments about each axis during the gait trials. From the 3 components of the GRF, the heart of pressure (COP), vertical torque and resultant GRF can exist computed. Effigy ane shows a participant pushing off from the plate during the late-stance phase of the gait wheel. Using the anterior-posterior shear force as an example, the force plate measures the force of the foot pushing back on the plate, which leads to a reaction force in the direction of travel.

During the belatedly stance phase of the gait cycle, the foot pushes downwardly and back on the force plate, causing equal and opposite reaction forces.

Figure 1.

Figure 1. During the late stance phase of the gait cycle, the foot pushes down and back on the force plate, causing equal and opposite reaction forces.

During the belatedly stance phase of the gait cycle, the foot pushes down and back on the strength plate, causing equal and reverse reaction forces.

Figure 1.

Figure 1. During the late stance phase of the gait cycle, the foot pushes down and back on the force plate, causing equal and opposite reaction forces.

During the late stance phase of the gait cycle, the foot pushes downwards and dorsum on the strength plate, causing equal and opposite reaction forces.

Close modal

Forcefulness plates usually consist of strain estimate or piezoelectric transducers and are available in a variety of sizes. For the purpose of gait analysis, these devices are embedded within the flooring or walkway of the laboratory to provide a level walking surface. Covering the forcefulness plates with flooring or carpet can forestall targeting of the plates and consequently, can prevent contradistinct joint kinematics and/or kinetics. The motion analysis laboratory at the University of New Brunswick employs a large strain gauge plate (BP5918, Advanced Mechanical Applied science Incorporated [AMTI], Newton, MA) and 4 smaller piezoelectric force plates (9281B11; 9281CA, Kistler Instruments, Winterthur, Switzerland). The dimensions of the AMTI plate are 1499 ×457 mm, with a natural frequency of 300 Hz. The 9281B11 and 9281CA Kistler plates are 600 ×400 mm, with natural frequencies of approximately 800 Hz and 1000 Hz, respectively. Using these plates, it is possible to obtain force plate data for a consummate left and right gait cycle in a single trial. Force plate information can exist sampled at frequencies equivalent to the camera organization (i.east. 50 Hz or 60 Hz), or force plate information can be sampled at higher frequencies and later subsampled to represent with the camera data.

The examination of the characteristics of the force curves generated during the gait wheel can provide valuable clinical data. For example, slow loading rates and/or apartment strength profiles during the gait bicycle may propose the presence of some pathology.11 Understanding the magnitude and the spatial relationship of the resultant GRF relative to specific joints during walking is as well important equally this influences the direction in which the articulation will tend to rotate.12 In addition, pathological gait may result in an altered center of pressure pattern or large twisting moment applied past the foot to the floor. To collect gait bicycle data for both lower extremities, at least two force plates are needed. Having three or four plates will increment the probability of a successful foot strike on the plate and decrease the number of trials the patient must perform. This is an of import issue if the patient suffers from fatigue and may be able to perform simply a express number of gait cycles.

Electromyography

Electromyography (EMG) refers to the recording of electrical activity produced during muscle contraction; the technology is used to determine the intensity of contraction and phasic action of the lower extremity muscles during normal and clinical gait.thirteen,14 Data from EMG provides insight into the underlying neuromuscular activity and aids in the estimation of the kinematic and kinetic information obtained from the aforementioned systems. From superficial muscles, EMG data tin can be obtained using surface electrodes placed on the patient'southward skin. A more direct and invasive method involves inserting fine wire electrodes into the muscle belly. This enables the measurement of specific and deeper muscle tissue and is less likely to suffer from cross-talk and tissue filter effects. However, this technique can cause pain and discomfort, which could alter the patients' movement patterns. Ideally, telemetered EMG systems are preferred for gait analysis, because they are less probable to interfere with the natural walking patterns of the patient.

Equipment Integration

In summary, gait laboratories commonly consist of 3 major systems of equipment: 1) a motion capture system, ii) strength plates, and 3) EMG. It is critical that these systems are precisely synchronized in society to: 1) relate the GRF data to the position of the body segments in space, and ii) relate the EMG muscle activeness to the joint move data. Motility capture software and a 64-aqueduct A/D lath (too as other analog devices) are capable of simultaneously collecting and synchronizing kinematic, kinetic, and EMG data. This profoundly simplifies the information collection process and required instrumentation, considering multiple triggering devices are not required.

The experimental data obtained from the motion capture system and force plates must be entered into a mechanical model to calculate parameters such as relative angles and articulation moments. The human body is generally modeled every bit a organization of rigid body segments with articulations of various degrees of freedom. Nigh mechanical models begin past defining body-fixed Cartesian coordinate systems for each rigid body using 3 noncollinear points (bony landmarks represented past the reflective markers). Figure ii depicts a typical marker set and the embedded coordinate systems on each segment. The origin of the coordinate systems for the ankle, human knee, hip, and pelvis are defined at the estimated joint centers. Joint centers can be estimated using a multifariousness of algorithms.fifteen,sixteen Modeling the body as a series of rigid segments facilitates the computation of kinematic and kinetic parameters at each of the joints.

Typical marking set and embedded coordinate axes. Blue circles represent reflective marking locations.

Effigy 2.

Figure 2. Typical marker set and embedded coordinate axes. Blue circles represent reflective marker locations. / Red circles represent estimated joint center locations and are the origin of the embedded coordinate systems (Figure reprinted with permission by Oxford Metrics Ltd).

Typical marker set and embedded coordinate axes. Blue circles represent cogitating marking locations.

Red circles represent estimated articulation middle locations and are the origin of the embedded coordinate systems (Figure reprinted with permission past Oxford Metrics Ltd).

Effigy 2.

Figure 2. Typical marker set and embedded coordinate axes. Blue circles represent reflective marker locations. / Red circles represent estimated joint center locations and are the origin of the embedded coordinate systems (Figure reprinted with permission by Oxford Metrics Ltd).

Typical marking set and embedded coordinate axes. Bluish circles stand for reflective marker locations.

Red circles represent estimated articulation center locations and are the origin of the embedded coordinate systems (Figure reprinted with permission by Oxford Metrics Ltd).

Close modal

Equations 1 to 7 provide calculations for the formation of a pelvic embedded coordinate system based on the three pelvic markers: left and right anterior superior iliac spine (lasi, rasi) and the sacrum (sacr). A virtual articulation was created midway between the rasi and lasi markers. This virtual joint represented the origin of the pelvic coordinate organization (Figure 2). A relative position vector from the sacral wand to the virtual joint was created and designated past r o/sacr (origin relative to sacr). This relative position vector was computed by subtracting the (x,y,z) marker coordinates in the post-obit manner:

formula

where r o equals the midpoint between the two anterior superior iliac spines:

formula

A 2d relative position vector was created from the correct to the left anterior superior iliac spine markers and was designated by r50/r

formula

A unit vector east y , was created along this transverse centrality past dividing the relative position vector past its magnitude (pelvic width):

formula

The third vector, r x , was computed from a Gram-Schmidt orthogonalization procedure,xv

formula

and so normalized to define unit vector due east 10 ,

formula

The final unit vector is computed from the cantankerous production of unit vectors e ten and e y :

formula

Kinematics

Kinematics is a branch of mechanics that describes motion without reference to the underlying crusade(southward) of the motion. Kinematic analyses of gait rely on the knowledge of the instantaneous position of each body segment. Given that the motion-capture system provides the 3-dimensional trajectories of the markers placed on the patient's peel, it is possible to decide the linear and angular displacement, velocity, and dispatch of the joints and body segments. These parameters are also required for kinetic analyses (meet Kinetics).

Temporal-Spatial Parameters

A gait cycle is defined equally the menstruum betwixt consecutive foot strikes of the aforementioned leg. Information technology is approximately i second in duration and can be subdivided into phases for analytical purposes. There are several different models of the gait bicycle that may be adopted;17 one such model is illustrated in Figure 3. The duration of time a person spends in a particular portion of the gait cycle is quantified and can provide valuable information about the stability and symmetry of motion patterns. For instance, patients suffering from instability (e.g. Parkinson's illness, hypotonia) may subtract walking speed and increment the percentage of the cycle spent in double support, where both feet are on the ground. Temporal-spatial variables such as cadence (steps/minute), walking velocity, stride length, and width of base of support tin also provide valuable data almost stability and move patterns.

Typical mature gait bike (Sutherland, 1981: The events of gait. Bulletin of Prosthetic Res. 1981;10[35];312–315. Figure reprinted with permission).

Effigy 3.

Figure 3. Typical mature gait cycle (Sutherland, 1981: The events of gait. Bulletin of Prosthetic Res. 1981;10[35];312–315. Figure reprinted with permission).

Typical mature gait wheel (Sutherland, 1981: The events of gait. Bulletin of Prosthetic Res. 1981;10[35];312–315. Figure reprinted with permission).

Effigy 3.

Figure 3. Typical mature gait cycle (Sutherland, 1981: The events of gait. Bulletin of Prosthetic Res. 1981;10[35];312–315. Figure reprinted with permission).

Typical mature gait cycle (Sutherland, 1981: The events of gait. Bulletin of Prosthetic Res. 1981;10[35];312–315. Figure reprinted with permission).

Close modal

Joint Angles

Joint angles, or relative segment angles, provide of import data on the spatial and temporal coordination between segments during the gait cycle. Joint rotation angles tin be obtained by calculating the relative orientation of the embedded coordinates systems in the proximal and distal segments. For example, knee rotations are rotations of the shank embedded coordinate organisation with respect to the thigh embedded coordinate organisation. Joint angle parameters have well-established normative values and are ane of the most frequently used kinematic measures.

The joint rotations tin can be expressed using several different methods, including helical axis, joint coordinate system, and Euler/Cardan angles. Many researchers and clinicians employ Euler angles to describe joint angles, which define angular motion as a sequence of ordered rotations nearly the axes of the Cartesian coordinate system.18,19 Although the rotations are sequence-dependent, they provide joint orientation in a manner that is physically meaningful and clinically relevant. For instance, the Euler angles of the hip are flexion/extension, adduction/abduction, and internal/external rotation. Detailed calculations of joint angles using embedded-coordinate systems are provided in Appendix A.

Kinetics

Kinetics is the branch of mechanics that is concerned with the forces and moments transmitted across the joints of the trunk. Move is facilitated by the interaction of passive and active forces acting on the body. To empathise the underlying causes of normal and clinical gait patterns, kinetic analyses must be performed. Kinetic parameters include joint moments and joint power. When treatments such as orthoses or surgical intervention are prescribed for a patient, it is these kinetic parameters that are influenced. Therefore, objective quantification and cess of these parameters is disquisitional for determining treatment effectiveness.

Joint moments are defined as the rotational potential of the forces interim on a articulation and are typically classified as external or internal in nature. Internal joint moments refer to the cyberspace outcome of all internal forces interim virtually a given joint, including those due to muscular contraction, soft tissues, friction, and joint configuration.12 External joint moments during gait generally refer to loads applied to the trunk segments due to basis reaction forces (e.g. every bit the human foot strikes the ground), inertial forces, and gravitational forces. In gait analyses, clinicians are attempting to judge the internal moments acting about the joints. The clinician is interested in identifying the type of movement being produced (i.e. flexor vs extensor moment) and the magnitude of the moment. The pattern and timing of the joint moment results are so examined and compared to normative data. For gait analysis, joint moments are by and large computed using the inverse dynamics approach.

Inverse Dynamics

Directly measurements of muscle forces and moments are not applied, because they require invasive procedures. A mathematical model, referred to as the changed dynamics approach, tin can be used to judge the three-dimensional net joint moments for the hip, knee joint, and ankle joints during gait.20,21 Changed dynamics requires the integration of motility data, force plate data, and inertia information. Using this approach, each body segment is modeled as a free torso. The components of the resultant force and resultant moment can exist calculated in a stepwise fashion, beginning with the distal foot and ending at the hip articulation. The forces and moments at the distal end of the foot are provided by the force platform data; however, the ankle joint forces and moments are unknown. Solving for these unknowns requires the add-on of linear and athwart acceleration information and segment inertial information. Linear and angular acceleration are obtained through numerical differentiation of the position and orientation data. The knee joint forces and moments are then calculated using the known ankle data from the previous calculations. The same procedure is repeated for the hip.

Newton-Euler equations of motion for dynamic equilibrium are used to solve the cyberspace joint reaction forces (F) and moments (M):

formula

formula

formula

formula

formula

formula

where

Fx, Fy, Fz = x y z components of the internet articulation reaction forcefulness,

M10, Thousandy, 1000z = x y z components of the internet joint moment about the center of mass (CM),

thou = the mass of segment i,

a CMx , a CMy , a CMz = x y z components of linear dispatch of the centre of mass of segment i,

α ten , α y , α z = x y z components of the angular acceleration of segment i,

I 20 , I yy , I zz = moment of inertia of segment i,

ω x , ω y , ω z = ten y z components of angular velocity of segment i,

All forces and moments are reported with respect to the axes of the embedded coordinate systems (Effigy 4). The use of Equations 11 through 13 assumes that the principal axes of inertia coincide with the master anatomic axes. Although McConville et al. (1980) reported that the orientation of these two axes systems could differ greatly, the differences in moments of inertia were small and similar to the mistake of the estimates.

Three-dimensional knee forces and moments are computed with respect to the embedded coordinate systems.

Figure iv.

Figure 4. Three-dimensional knee forces and moments are computed with respect to the embedded coordinate systems.

Three-dimensional knee forces and moments are computed with respect to the embedded coordinate systems.

Figure 4.

Figure 4. Three-dimensional knee forces and moments are computed with respect to the embedded coordinate systems.

Three-dimensional knee forces and moments are computed with respect to the embedded coordinate systems.

Shut modal

The inverse dynamics approach is oftentimes used in gait assay, considering it is relatively accurate, unproblematic, and noninvasive. However, it is important to emphasize that the results of this model stand for the internet joint moments during walking. As a outcome, this approach likely underestimates the loading at the joint. For instance, persons with cerebral palsy frequently prove simultaneous agonist and adversary musculus activity (cocontraction) most the joints. If cocontraction of the hamstrings and quadriceps occurred at the hip joint, inverse dynamics would not report the individual muscle force contribution, but rather the cyberspace result of the two musculus groups. For this reason, the results should be accompanied past EMG data to provide information on the phasic action and intensity of muscle contractions.

A greater understanding of the role of muscles in producing and decision-making motion tin be achieved by calculating joint ability. Articulation power is typically defined as the production of joint moment and articulation angular velocity. Positive ability refers to the generation of free energy through concentric wrinkle of the muscle, whereas negative power refers to the absorption of energy through an eccentric wrinkle. This definition of joint power assumes that the translation of ane segment with respect to another is negligible (three degrees of freedom). More complex definitions of joint power (P) incorporate both the rotational and translational powers using the post-obit equation:

formula

where M J is the joint moment, ω J is the joint angular velocity, F J is the internet joint force, and 5 J is the joint translational velocity. Inquiry suggests that this six-degree of freedom model is a more authentic estimate of articulation ability.22,23

Segment Inertia Model

Segment inertia refers to the mass and distribution of mass inside a segment. To calculate joint moments during walking, the size and inertial properties of the lower limb segments are required (see Equations 8–13). Average adult segment inertial data are well documented24–27 and several researchers accept also provided data for infants and children.28–thirty These latter estimates are not always applicable to children with developmental disorders or torso segment deformities. Inertial parameters for these populations tin exist approximated using mathematical models of the human being body, such as the elliptical cylinder method.31 The elliptical cylinder method is currently employed at the University of New Brunswick to obtain child segment inertial values. This technique is ideal for clinical populations who may have limb or segment deformities, because the method provides estimates of segment inertia based on each child'south body paradigm. The technique requires that digital front and side images of each patient exist recorded while he or she is continuing in the anatomical position. These total-torso images are and then outlined and manually digitized. The model consists of 16 segments and each segment is assumed to consist of elliptical cylinders created at 1- to 2-cm intervals in the transverse aeroplane (Effigy 5). Given that the volume and density of each elliptical cylinder is known, the mass of each elliptical cylinder tin can be calculated. The segment mass, center of mass location, and moments of inertia are then calculated from the stacked elliptical cylinders representing each segment. Compared with methods that apply average values, this mathematical model provides individually tailored estimates of segment inertial parameters. Therefore, it is platonic for clinical applications, because it is able to provide estimates for individuals who do not comply with average values (east.g. amputees, individuals with deformities).

Elliptical cylinder model of a 4-twelvemonth-old male child.

Effigy v.

Figure 5. Elliptical cylinder model of a 4-year-old boy.

Elliptical cylinder model of a 4-twelvemonth-onetime boy.

Figure five.

Figure 5. Elliptical cylinder model of a 4-year-old boy.

Elliptical cylinder model of a 4-year-old boy.

Close modal

I of the primary objectives of the interpretation process is to identify deviations in a patient'south gait from normal movement patterns, and to establish the underlying causes of these differences. Once the causes of the deviations accept been determined, it may be possible to formulate treatment plans to improve the functional status of the patient.four Handling may consist of physiotherapy, surgical intervention, medication, and/or orthoses. Information technology is critical for the clinician to identify any gait adaptations that may be present to ensure the root of the problem is being treated, and not the compensation. Additional gait analyses tin and so be performed to evaluate movement patterns with and without treatment (due east.m. walking with and without an orthosis), pre-and postsurgery, and patient progress.

Comprehensive gait analyses produce big numbers of variables and the interpretation of this data tin be very complex and time-consuming. As a consequence, many clinicians continue to use subjective methods of gait evaluation. To facilitate the utilise of gait analyses in the clinical setting, the large volumes of data must be reduced and presented in a manner that is both practical and interpretable. Researchers at the University of New Brunswick have created statistical models that automatically reduce data by measuring a patient'due south variability nearly mean normative values for sagittal joint angle data.32 The statistical model computes a 1-dimensional score of normality that reflects the extent of the difference between the patient and normative data. If a patient receives a high score, suggesting abnormal movement patterns, the clinician is able to quickly identify which sagittal joint bending is aberrant through the scoring algorithm. Equally near gait variables are continuous, the model is as well capable of identifying when the abnormality occurs in the gait bike. For example, a patient may have normal knee flexion during periods of double support, even so show abnormal patterns of flexion during single stance. The data reduction method will place the variable "knee flexion" equally deviant and identify the portion of the gait wheel in which the difference occurred. Currently, this statistical model is being expanded to incorporate kinetic variables and boosted kinematic variables.10

Clinical gait analysis is a powerful tool to appraise abnormal gait patterns, treatment effectiveness, and the furnishings of disease progression. Current research at the University of New Brunswick examines gait patterns and treatment effectiveness in children with hypotonia and spastic diplegia. Built hypotonia, which is a symptom of a number of developmental disorders (i.e. Downwards syndrome), is characterized past the delayed onset of independent walking, aberrant walking patterns, and lower limb deformities. Typical treatments include casting, orthotics, and physical therapy.33 Extensive orthotic treatment of this population is relatively new and somewhat controversial. Subjective physiotherapy reports advise that these treatments result in an improved walking ability. To appointment, in that location are few objective studies that address the effectiveness of these treatments.34,35 We examined the effectiveness of ankle-foot orthoses in hypotonic children, aged i to thirteen years. Participants were asked to walk with and without their orthoses in the gait lab. Barefoot walking patterns were compared to age-matched normative data and orthosis trials.

The sagittal joint angle results for 1 participant are shown in Figure 6. To compensate for weak calf muscles, which allow the tibia to accelerate rapidly over the planted foot, this hypotonic participant has opted to strength the knee back into hyperextension during stance stage. This reduces the demand on the quadriceps and stabilizes the lower extremity. When walking with an orthosis, the participant shows increased hip and human knee flexion and ankle dorsiflexion, compared to barefoot gait. The results of the orthosis trials approximate the normative data more closely than the results of barefoot trials exercise. Figure 7 shows the improved sagittal ankle moment as a function of the orthosis. Without the orthosis, the participant showed relatively little plantarflexor moment, which was likely due to astringent pronation and pain during weightbearing. The orthosis controlled the alignment of the ankle joint, reduced hurting, and increased joint stability, which facilitated improved gait patterns.

Sagittal hip, articulatio genus, and ankle angles for a child with hypotonia, walking with orthoses (o) and barefoot (+).

Figure 6.

Figure 6. Sagittal hip, knee, and ankle angles for a child with hypotonia, walking with orthoses (o) and barefoot (+). / Mean normative data (± 1 SD) is shown for reference.

Sagittal hip, knee, and ankle angles for a kid with hypotonia, walking with orthoses (o) and barefoot (+).

Mean normative data (± 1 SD) is shown for reference.

Figure 6.

Figure 6. Sagittal hip, knee, and ankle angles for a child with hypotonia, walking with orthoses (o) and barefoot (+). / Mean normative data (± 1 SD) is shown for reference.

Sagittal hip, articulatio genus, and ankle angles for a child with hypotonia, walking with orthoses (o) and barefoot (+).

Hateful normative data (± one SD) is shown for reference.

Close modal

Sagittal ankle moment for a child with hypoto-nia, walking with orthoses (o) and barefoot (+).

Figure 7.

Figure 7. Sagittal ankle moment for a child with hypoto-nia, walking with orthoses (o) and barefoot (+). / Mean normative data (± 1 SD) is shown for reference.

Sagittal talocrural joint moment for a child with hypoto-nia, walking with orthoses (o) and barefoot (+).

Mean normative information (± ane SD) is shown for reference.

Figure seven.

Figure 7. Sagittal ankle moment for a child with hypoto-nia, walking with orthoses (o) and barefoot (+). / Mean normative data (± 1 SD) is shown for reference.

Sagittal ankle moment for a kid with hypoto-nia, walking with orthoses (o) and barefoot (+).

Mean normative data (± i SD) is shown for reference.

Close modal

We are also examining the effectiveness of series casting in children with spastic cerebral palsy (CP). Cognitive palsy is a chronic, nonprogressive disorder that mainly affects gait and posture. Serial casting is an aggressive but noninvasive therapy that involves repeatedly casting joints in an elongated position to increment range of motion and muscle fiber length, and to decrease spasticity and joint contractures. Subjectively, serial casting has proven to exist an constructive treatment modality for improving motor ability in children with CP. In addition, serial casting is an fantabulous alternative to surgery in growing children. To assess treatment success, physiotherapists typically utilise passive range of motility (ROM) tests and subjective gait evaluations. These subjective measures of walking are limited in that they are, to a swell extent, dependent on the clinician's experience and visual memory. In improver, the data obtained cannot be readily quantified, which leads to difficulty when the clinician tries to demonstrate improvements pre- and post-treatment. Therefore, objective evaluations of the effectiveness of serial casting on gait patterns of children with CP are currently being conducted. Gait information are collected pre- and mail service-treatment and compared to age-matched normative data. The results are compared to typical physiotherapy evaluations. Of interest, is whether the passive ROM tests and subjective gait evaluations will exist correlated with the results from the objective gait analyses.

Objective gait analyses have provided clinicians with a powerful tool to evaluate treatment interventions and to increase our agreement of normal and clinical gait patterns. Recent advances in computer technology and the application of mechanical engineering accept facilitated the rapid progress of the field of gait assay. Objective gait analyses provide valuable information for orthotic design and surgical and treatment decisions. Knowledge of which handling or combination of treatments is about effective will lead to more successful handling interventions, improved functional status for the patient, and reduced health care costs. Current research efforts should be directed towards improved analytical and data reduction techniques to increase the viability of objective gait analyses in the clinical setting.

The authors wish to gratefully acknowledge the Natural Sciences and Technology Research Council (NSERC), the New Brunswick Women'due south Doctoral Scholarship Programme, and the Establish of Biomedical Engineering, University of New Brunswick, Fredericton, New Brunswick, Canada.

Kaufman

,

Thousand. R.

1998

.

Gait Analysis in the Science of Rehabilitation

.

Vankoski

,

Due south.

and

L.

Dias

.

Clinical motility analysis.

J Child Mem Hosp

1999

.

Spring Ed

.

Inman

,

V. T.

,

H. J.

Ralston

, and

F.

Todd

.

Human being Walking

.

Baltimore, MD

Williams & Wilkins

.

1981

.

Davis

,

R. B.

Reflections on clinical gait analysis.

J Electromyogr Kinesiol

1997

.

7

4

:

251

257

.

Elftman

,

H.

The measurement of the external force during walking.

Science

1938

.

88

:

152

153

.

Bresler

,

B.

and

J. P.

Frankel

.

The forces and moments in the leg during level walking.

Am Soc Mech Eng Trans

1950

.

72

:

27

36

.

Whittle

,

M. West.

Clinical gait analysis: a review.

Hum Mov Sci

1996

.

15

:

369

387

.

Vicon 512 Movement Capture System. Oxford Metrics Ltd

.

Cappozzo

,

A.

,

F.

Catani

,

A.

Leardini

, et al.

Position and orientation in space of bones during movement: experimental artifacts.

Clin Biomech

1996

.

11

2

:

90

100

.

Chester

,

V. 50.

Identification of clinically relevant variables in hypotonic gait

[unpublished dissertation]. Fredericton, NB, Canada: University of New Brunswick

.

2003

.

Biden

,

Eastward.

,

J.

O'Connor

, and

J. J.

Collins

.

Gait Analysis.

In:

Daniel

,

D.

, et al

Knee joint Ligaments: Structure, Role, Injury and Repair

.

New York

Raven Press

.

1990

.

Ayyappa

,

E.

Normal homo locomotion, office 2: motion, footing reaction strength and muscle activity.

J Prosthet Orthot

.

1997

.

ix

2

:

42

57

.

Sutherland

,

D. H.

The development of clinical gait assay function 1: kinesiological EMG.

Gait Posture

2001

.

xiv

:

61

lxx

.

Perry

,

J.

Gait Analysis: Normal and Pathological Function

.

Thorofare, NJ

Slack

.

1992

.

Davis

,

R. B.

,

S.

Ounpuu

,

D.

Tyburski

, and

J. R.

Gage

.

A gait analysis and information collection and reduction technique.

Hum Mov Sci

1991

.

10

:

575

587

.

Bong

,

A. L.

,

R. A.

Make

, and

D. R.

Pederson

.

Prediction of hip joint center locations from external landmarks.

Hum Mov Sci

1989

.

viii

:

3

16

.

Thomas

,

Due south. S.

and

T. J.

Supan

.

A comparing of current biomechanical terms.

J Prosthet Orthot

1990

.

ii

two

:

107

114

.

Chao

,

E. Y. S.

Justification of tri-centric goniometer for the measurement of joint rotation.

J Biomech

1980

.

13

:

989

1006

.

Grood

,

E. S.

and

W. J.

Suntay

.

A joint coordinate organization for the clinical description of three-dimensional motions: application to the knee.

J Biomech Eng

1983

.

105

:

136

144

.

Wittenburg

,

J.

Dynamics of Systems of Rigid Bodies

.

Stuttgart, Germany

Teuber Verlag

.

1977

.

Õunpuu

,

S.

,

R. B.

Davis

, and

P. A.

DeLuca

.

Joint kinetics: methods, interpretation and treatment decision-making in children with cerebral palsy and myelomeningocele.

Gait Posture

1996

.

4

:

62

78

.

Duncan

,

J. A.

,

D. Fifty.

Kowalk

, and

C. L.

Vaughan

.

Six degree of liberty joint power in stair climbing.

Gait Posture

1997

.

5

:

204

210

.

Buczek

,

F. L.

,

T. M.

Kepple

,

Chiliad. Fifty.

Siegel

, et al.

Translational and rotational joint power terms in a six caste-of-freedom model of the normal talocrural joint complex.

J Biomech

1994

.

27

12

:

1447

1457

.

Dempster

,

W. T.

Space Requirements of the Seated Operator

.

Wright-Patterson Air Force Base of operations

.

Ohio

.

1955

.

WADC Technical Report 55–159

.

Clauser

,

C. Due east.

,

J. T.

McConville

, and

J. West.

Young

.

Weight, Volume, and Center of Mass of Segments of the Human Body

.

Wright-Patterson Air Forcefulness Base

.

Ohio

.

1969

.

Technical Study AMRL-TDR-69-70

.

Chandler

,

R. F.

,

C. E.

Clauser

,

J. T.

McConville

, et al.

Investigation of the Inertial Properties of the Human Body

.

Wright-Patterson Air Force Base

.

Ohio

.

1975

.

Technical Written report AMRL-TR-74-37, Advertizing-A016-485, DOT-HS-801-430

.

Zatsiorsky

,

V. G.

and

V. Due north.

Seluyanov

.

The mass and inertia characteristics of the principal segments of the body.

In:

Matsui

,

H.

and

K.

Kobayashi

.

Biomechanics Viii-B

.

Champaign, IL

Man Kinetics

.

1983

. :

1152

1159

.

Sun

,

H.

and

R. Thou.

Jensen

.

Body segment growth during infancy.

J Biomech

1994

.

27

3

:

265

275

.

Schneider

,

K.

and

R. F.

Zernicke

.

Mass, heart of mass, and moment of inertia estimates for infant limb segments.

J Biomech

1992

.

25

2

:

145

148

.

Jensen

,

R. K.

Trunk segment mass, radius, and radius of gyration proportions for children.

J Biomech

1986

.

19

5

:

359

368

.

Jensen

,

R. K.

Estimation of the biomechanical properties of iii body types using a photogrammetric method.

J Biomech

1978

.

11

:

349

358

.

Tingley

,

M.

,

C.

Wilson

,

Due east.

Biden

, and

West. R.

Knight

.

An index to quantify normality of gait in young children.

Gait Posture

2002

.

16

:

149

158

.

Caselli

,

M. A.

,

E.

Cohen-Sobel

,

J.

Thompson

,

J.

Adler

, and

L.

Gonzalez

.

Biomechanical direction of children and adolescents with Down's syndrome.

J Am Podiatr Med Assoc

1991

.

81

3

:

119

127

.

Selby-Silverstein

,

L.

,

Southward.

Bhimji

,

H. J.

Hillstrom

, et al.

Moment about the lower leg during gait of children with Downwardly syndrome and the effect of foot orthoses.

Gait Posture

1995

.

three

ii

:

97

..

Selby-Silverstein

,

L.

,

H. J.

Hillstrom

, and

R. J.

Palisano

.

The effect of foot orthoses on standing human foot posture and gait of young children with Down's syndrome.

Neuro Rehabilitation

2001

.

sixteen

3

:

183

93

.

Appendix

To compute joint angles, embedded coordinate systems are created at the estimated joint centers. Joint rotation angles can then be obtained by calculating the relative orientation of the embedded coordinates systems in the proximal and distal segments. For instance, knee rotations are rotations of the shank embedded coordinate system with respect to the thigh embedded coordinate system. The relation of the two embedded coordinate systems in the body segments are related to 1 another past the 3 contained angles which tin be mathematically defined past:

formula

where:

[R zxy ] = rotation matrix for an lodge of rotations

[R 10 x )] = rotation matrix for a rotation (Φ 10 ), about the axis x (same for y and z)

If θ y , θ x , and θ z are planar rotations virtually the x, y, and z axes of a Cartesian coordinate system, the component rotations become:

formula

formula

formula

One set of solutions for the three angles is:

formula